FLUKE DIGITAL MULTI METER MODEL 106+
Download Fluke 106 Palm-sized Digital Multimeter | Data sheets
Voltage, resistance, continuity, capacitance
Input terminal for AC and DC current measurements to 10 A and current
Data hold
Specifications: Fluke 106 Palm-sized Digital Multimeter
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits])
Accuracy Specifications | ||||
Function | Range | Resolution | Accuracy | |
AC Volts (40 Hz to 500Hz)1 | 6.000 V 60.00 V 600.0 V | 0.001 V 0.01 V 0.1 V | 1.0 % + 3 | |
DC Volts | 6.000 V 60.00 V 600.0 V | 0.001 V 0.01 V 0.1 V | 0.5 % + 3 | |
AC Millivolts | 600.0 mV | 0.1 mV | 3.0 % + 3 | |
Diode Test2 | 2.000 V | 0.001 V | 10 % | |
Resistance (Ohms) | 400.0 Ω 4.000 kΩ 40.00 kΩ 400.0 kΩ 4.000 MΩ 40.00 MΩ | 0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ | 0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5% + 3 | |
Capacitance3 | 50.00 nF 500.0 nF 5.000 μF 50.00 μF 500.0 μF 1000 μF | 0.01 nF 0.1 nF 0.001 μF 0.01 μF 0.1 μF 1 μF | 2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 | |
Frequency4 Hz (10 Hz – 100 kHz) | 50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz | 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz | NA | |
Duty Cycle4 | 1 % to 99 % | 0.1 % | NA | |
AC Current (40 Hz to 200 Hz) | 4.000 A 10.00 A | 0.001 A 0.01 A | 1.5 % + 3 | |
DC Current | 4.000 A 10.00 A | 0.001 A 0.01 A | 1.5 % + 3 | |
| ||||
Function | Overload Protection | Input Impedance (Nominal) | Common Mode Rejection Ratio | Normal Mode Rejection Ratio |
AC Volts | 600 V 1 | >10 MΩ <100 pF 2 | >60 dB at dc, 50 Hz or 60 Hz | − |
\AC Millivolts | 600 mV | >1M, <100 pF | >80 dB at 50 Hz or 60 Hz | − |
DC Volts | 600 V 1 | >10 MΩ <100 pF | >100 dB at dc, 50 Hz or 60 Hz | >60 dB at 50 Hz or 60 Hz |
| ||||
General Specifications | ||||
Maximum Voltage Between any Terminal and Earth Ground | 600 V | |||
Display (LCD) | 6000 counts, updates 3/sec | |||
Battery Type | 2 AAA, NEDA 24A, IEC LR03 | |||
Battery Life | 200 hours minimum | |||
Temperature | ||||
Operating | 0 °C to 40 °C | |||
Storage | -30 °C to 60 °C | |||
Relative Humidity | ||||
Operating Humidity | Non-condensing when <10°C ≤90 % at 10 °C to 30 °C; ≤75 % at 30 °C to 40 °C | |||
Operating Humidity, 40 MΩ Range | ≤80 % at 10 °C to 30 °C; ≤70 % at 30 °C to 40 °C | |||
Altitude | ||||
Operating | 2000 m | |||
Storage | 12,000 m | |||
Temperature Coefficient | 0.1 X (specified accuracy) /°C (<18 °C or >28 °C) | |||
Fuse Protection for Current Inputs | 11A, 1000V Fast Fuse, Fluke specified part only | |||
Size (HxWxL) | 142 mm x 69 mm x 28mm | |||
Weight | 200 g | |||
IP Rating | IEC 60529: IP 40 | |||
Safety | IEC 61010-1: 600 V CAT III, Pollution Degree 2 | |||
Electromagnetic Environment | IEC 61326-1: Portable | |||
Electromagnetic Compatibility | Applies to use the Korea only. Class A Equipment (Industrial Broadcasting and Communication Equipment)1 | |||
|